Sensitive, Flexible Workhorse that Makes Assay Development Almost Painless

NTHU
IMCB
Graduate Student

Overall

Performance

Ease-of-Optimization

What do these ratings mean?
Write a Review

Company:

BMG Labtech

Product Name:

BMG CLARIOstar plus plate reader

Catalog Number:

The BMG LABTECH CLARIOstar Plus is a multi-mode plate reader for [e.g., ELISA, fluorescent viability assays, luciferase reporters, and TR-FRET], and it has become the instrument I trust most for both routine plates and new assay development. Continuously adjustable bandwidths make it easy to match unusual fluorophores or optimize signal-to-background for a new assay.

Experimental Design and Results Summary

Performance, Range, and Accuracy

The CLARIOstar Plus covers absorbance, FI/FRET, luminescence/BRET, FP, TRF/TR-FRET, and optional Alpha on 6- to 1536-well plates, with EDR giving ~8 decades of range without manual gain. Absorbance (220–1000 nm) is accurate to <1% at 2 OD, and fluorescence/luminescence LODs are in the sub-pM to fM range depending on mode. Autofocus and spectral scanning make it reliably hit the intended detection window for both routine plates and new assays.

Throughput, Workflow Efficiency, and Automation

A 96-well plate can be read in ~8 s (flying mode); 384-well in ~15 s; full absorbance spectra take <1 s/well. LVF monochromators, EDR, optional injectors, and a 50-plate stacker cut setup and hands-on time, with robot/LIMS control via ActiveX, DDE, and ASCII. Without a stacker it is still one-plate-at-a-time loading, so the biggest labor saving is in assay development rather than ultra-HTS campaigns.

Sample Integrity and Consumable Requirements

It safely reads standard SBS plates, live-cell assays (especially with ACU/bottom reading), and 2 µL drops on the LVis Plate; corrosive samples should not be left in the chamber. There is no per-read kit—ongoing costs are plates, reagents, optional gases/filters/injector parts, and periodic QC plates. Software licenses are unlimited, so day-to-day running cost is low relative to the capital purchase.

Footprint, Facilities, and Daily Maintenance

The unit is compact , runs on standard 100–230 V , and needs 15–35 °C, non-condensing humidity, plus optional CO₂/O₂ if the ACU is installed. Daily care is wiping the plate carrier, flushing injectors after use, and letting temperature/atmosphere equilibrate before kinetics. Annual PM and an Evaluation Plate for IQ/OQ are enough for most labs.

Connectivity, Data Output, and Software Compliance

Voyager/SMART Control acquire data and MARS analyzes it; files export to Excel, ASCII, and databases for LIMS and robotics. SMART Control and MARS are designed to meet FDA 21 CFR Part 11, with Evaluation/LVis plates supporting IQ/OQ and NIST-traceable checks. The reader is research-use-only, so regulated labs still need their own validation, user accounts, and backups.

Image Gallery

Summary

The Good

Excellent sensitivity and assay-development flexibility with almost no daily consumable burden

The Bad

A premium price and optional modules you must buy if you need stacker, injectors, or live-cell gas control

The Bottom Line

The CLARIOstar Plus is a highly capable, easy-to-run multi-mode workhorse that reliably covers most absorbance, fluorescence, luminescence, and TR-FRET needs in one compact instrument. It earns its cost in shared or core labs that develop and switch assays often; a simpler, cheaper reader is enough if you only run one endpoint all day. For research use, I would recommend it.

Join the discussion