CD19 on a Red Laser Color

Pathology
UMass
Lab Technician

Overall

Quality of Results

Ease-of-Optimization

What do these ratings mean?
Write a Review

Company:

Beckman Coulter

Product Name:

CD19-APC-A750, J3-119, 0.5 mL, ASR

Catalog Number:

A78838

Clone Number:

J3-119

Info: Supplier Page View Product Specs

Browse Similar Products: CD19 Antibodies

In research, it is often important to "reserve" common channels for unusual antibodies. Therefore, it is great that common antibodies, eg against CD19, can be found conjugated to fluorochromes that aren't used on daily basis. We needed CD19 on one such color, and we got it on APC-A750 (the antibody in review).

Experimental Design and Results Summary

Applications

Flow Cytometry

Sample

Whole blood (100uL)

Primary Incubation

15-30 minutes at room temp

Blocking Agent

1% FBS

Secondary Incubation

N/A

Tertiary Incubation

N/A

Detection

Beckman Navios flow cytometer

Results Summary

The antibody produced a good signal, showing reasonable resolution of the positive B cells from the negative, T and NK lymphocytes. Being a red-laser dye, there was a lot of compensation needed in both the APC and the APC-A700 channels. This is something that all labs need to be aware of. Otherwise, the staining was easy and the signal was good.

DOI or PMID #

Additional Notes

Related Categories

Image Gallery

Summary

The Good

Good signal. ASR.

The Bad

Expensive. Lots of compensation needed.

The Bottom Line

The antibody performed well, requiring little optimization but lots of compensation. The ASR tag is preferable in clinical trials.