Senescence is a driver of age-related diseases. Our research focuses on therapies targeting senescent cells and tissues. One of the easiest and fastest ways to induce senescence in vitro cell cultures is through irradiation.

Experimental Design and Results Summary

Application

Irradiation of cell cultures to induce premature senescence

Tips

The equipment has several drawers at different distances from the laser. Even though the same intensity of irradiation can be achieved using the different drawers (using different irradiation periods), it is important to keep using the same drawer/distance from the laser throughout the study to induce senescence in the same cell line. This because, to achieve the same irradiation intensities at different distances from the laser, cells will be irradiated for different time periods, which might influence the internal pathways of senescence induction.

DOI or PMID #

N/A

Related Categories

Image Gallery

Summary

The Good

User-friendly, easy and quick to use. Can be used for both cells and small rodents.

The Bad

The equipment has a substantial size. Not easy to transport to other rooms or facilities.

The Bottom Line

This irradiator is easy to use and requires no extra security or shielded room. It is self-contained and safe to use in a normal facility. Can be used for the irradiation of small rodents, such as mice, and also for cells. It is easy to adjust the irradiation intensities. Overall, this equipment is very user-friendly and requires little training to be used.