Detecting CD56

Toxicology
University of Kentucky
Postdoctoral Fellow

Overall

Quality of Results

Ease-of-Optimization

What do these ratings mean?
Write a Review

Company:

BD Biosciences

Product Name:

CD56 PE-Cy™7

Catalog Number:

335809

We do a lot of NK cell detection. We wanted to test a few CD56 antibodies and find the "best" one which would give us the easiest resolution of NK cells from other lymphocytes (see my other reviews). This product was chosen because PE-Cy7 is a bright fluorochrome and is available on our cytometer.

Experimental Design and Results Summary

Applications

Flow Cytometry

Sample

Human WBC

Primary Incubation

Using 5uL of the antibody for 15 minutes, 4C in the dark

Blocking Agent

None

Secondary Incubation

None

Tertiary Incubation

None

Detection

Flow Cytometer (PE-Cy7 channel)

Results Summary

The resolution of NK cells (shown in purple) from other lymphocytes was OK, but not great. We also saw quite a lot of ?non-specific positive signal on monocytes (shown in green) which made it harder to resolve the NK cells.

DOI or PMID #

N/A

Additional Notes

N/A

Related Categories

Image Gallery

Summary

The Good

An option for adding CD56 in a congested panel

The Bad

Resolution not very good. Monocytes appear as if positive

The Bottom Line

This antibody needs to be used with caution. Make sure you do not include monocytes in your NK gate because they can appear CD56-positive!

Share your experience with other scientists. Write a Review! »

Join the discussion