The C-digit Blot Made High Sensitivity Visualization Much Easier

Overall

Performance

Ease-of-Optimization

What do these ratings mean?
Write a Review
Hawaii Pacific University
College of Natural and Computational Sciences
Graduate student, instructor, and research assistant.

Company:

LI-COR

Product Name:

C-Digit Blot scanner

Catalog Number:

3600-00

We measure low abundance protein biomarkers in partially degraded marine mammal samples utilizing western blot, and rapid high sensitivity detection is vital. We selected this product for high value for money it offers, its compact side and ease of use.

Experimental Design and Results Summary

Tips

Wash membranes well to prevent background signal. Use a luminescent ladder or mark the ladder with a luminescent pen before putting in chemilum substrate. Update the scanner software regularly. Place a few drops of the chemilum solution on the scanner surface under the membrane. Balance the scanner to prevent the membrane from slipping during the scan. Never let the membrane dry out before scanning.

Related Categories

Image Gallery

Summary

The Good

Ease of use, relatively low cost, high sensitivity makes it good for low abundance protein visualization.

The Bad

Only one membrane can be scanned at a time, image production cannot be monitored during the 12 minute scanning process.

The Bottom Line

We are very happy with the performance of the C-digit blot, the ease of use, value for money, and technical support given by LICOR.