Fluorescence Readings For Small Samples

University of Colorado, Denver
School of Pharmacy
Post Doc

Overall

Performance

Ease-of-Optimization

What do these ratings mean?
Write a Review

Company:

Molecular Devices

Product Name:

SpectraMax M5 Multi-Mode Microplate Reader

Catalog Number:

SpectraMax M5

The general aim of the experiment was to generate standard curve of Nile red dye in DCM. We wanted to see the concentration of nile red within linear limit of fluorescence so as to quantify the dye in sample of interest. We selected the instrument because of its sensitivity and accuracy. We previously have noticed that this instrument can detect very low fluorescence in the order of 1 RFU or less that too with very small volume of sample (200ul).

Experimental Design and Results Summary

Application

Generation of standard cruve

Starting Material

1mg/ml nile red in DCM

Protocol Overview

The dye used was obtained from Sigma and was of extra purity. The solvent used was also of extra purity.

Tips

Take the blank reading in between of every reading.

Results Summary

We observed that upto around 100 ug/ml and less the curve was linear, but above that concentration the signal starts to saturate. Hence this curve can be used to quantify the dye from unknown sample if the signal intensity falls in linear range. Otherwise it is recommended to dilute the sample accordingly to bring the signal intensity within linear range.

Features Summary

Can measure multiple excitation emission wavelength simultaneouly

DOI or PMID #

N/A

Additional Notes

N/A

Image Gallery

Summary

The Good

Less samples, high sensitivity.

The Bad

Not applicable

The Bottom Line

This can be used to estimate the amount of any fluorescent material in the sample based on standard curve.