1uL Lag-3-PCPeF710 Works Great

Rheumatology
Vanderbilt University
Postdoctoral Fellow

Overall

Quality of Results

Ease-of-Optimization

What do these ratings mean?
Write a Review

Company:

eBiosciences

Product Name:

Anti-Human CD223 (LAG-3) PerCP-eFluor® 710

Catalog Number:

46-2239

Our group studies T cells in the blood of untreated chronic HIV+ individuals. We are constantly expanding and optimizing our T cell panels to measure various T cell subsets, activation, and inhibitory receptor phenotype. We picked Lag-3-PCPeF710 because the channel was open in a panel of 13 colors focused on inhibitory receptor expression on lymphocytes.

Experimental Design and Results Summary

Applications

Flow Cytometry

Sample

Human PBMC

Primary Incubation

1uL Lag-3-PCPeF710 in 200uL PBS with 10^6 cells; 15 minutes; room temperature

Blocking Agent

None

Secondary Incubation

None

Tertiary Incubation

None

Detection

BD LSRII Flow Cytometer

Results Summary

1uL Lag-3-PCPeF710 worked great in a 13-color panel to detect Lag-3+ cells. As expected, Lag-3 expression was low without stimulation but higher after overnight stimulation with SEB (1ug/mL). Of note, we had CD3 on the AF700 channel and repeated FMOs indicated there was no bleedover between the two channels at the titrated volumes chosen for each antibody.

Additional Notes

None

Related Categories

Image Gallery

Summary

The Good

1uL Lag-3-PCPeF710 works great; good option for big multi-color panel!

The Bad

None.

The Bottom Line

Lag-3-PCPeF710 works great to detect Lag-3+ lymphocytes in human PBMC.