advertisement
Advertisement (image not found)

Trouble-shooting: Low Yield

ShareThis

Trouble-shooting: Low yield

Here are some troubleshooting hints that we have gathered regarding low yield after PCR reactions:
 
  bullet2.gif (329 Byte) Primer annealing temperature is too high
Template not clean or degraded
  An enzyme inhibitor is present in the reaction
Not enough template in the reaction
Extension temperature too high
Enzyme activity is low
Cycle number is too high
Nucleotides hydrolized
Nucleotide concentration is too high or unbalanced
Primer concentration is too low
Machine based error
Plateau effect
Evaporation
 
 
Primer annealing temperature is too high

Primer annealing temperature is typically 50-60°C (may be higher or lower based on primer sequence and buffer components).

Suggestion:

Determine Tm/annealing temperature based on one of the following equations:

If primers are 20-35 bases If primers are 14 - 70 bases
Tp = 22 + 1.46(Ln)
Ln = 2(# G or C) + (# A or T)
TP = Effective annealing
temperature ± 2 - 5
Tm = 81.5 + 16.6 (log10 [J+]) + 0.41
(% G + C) - (600/l) - 0.063
(% Formamide) + 3 to 12
[J+] = concentration of Monovalent cations
l = length of oligo
 
Back to Top
 
 
Template not clean or degraded

For example, protease contamination can degrade the polymerase.

Suggestions:

  1. Try "Hot start" technique
  2. Purify the DNA template to the highest degree possible, including proteinase K digest
Back to Top
 
 
An enzyme inhibitor is present in the reaction
 

Suggestion:

Reduce or remove the concentration of any inhibitor in the reaction mixture. Known inhibitors of PCR include: 50 mM ammonium chloride
EDTA (metal chelator)
> 0.8 µM Hematin
PBS (phosphate will bind free magnesium)
> 0.02 % sarcosyl
0.5 M urea
> 5 % DMF
> 10 % formamide
heparin
> 20 % PEG deoxycholate
> 0.01 % SDS (can be reversed with equal molar ratio of NP40 and Tween 20)
> 10 % DMSO
> 20 % glycerol
> 0.4% N-Octylglucoside
Residual Phenol
>0.06% Sodium deoxycholate

Back to Top

 
 
Not enough template in the reaction

The necessary amount of template varies from reaction to reaction. As a guideline, 100 - 750 ng human DNA (105 - 106 copies) per 100 µl reaction. The amount of enzyme should be optimized for each template.

Suggestions:

  1. Titrate the amount of template in the reaction
  2. Perform optimization experiment varying enzyme concentration by 0.50 increments in suggested range (0.5 to 5.0 units)
Back to Top
 
 
Extension temperature too high
Optimal extension temperature and time is dependent of frament size:
  • 72°C for 20 seconds for fragments < 500 bp
  • 72°C for 40 seconds for fragments 1.2 bp.

Suggestion:

Longer times, not higher temperatures should be used when longer templates or suspected secondary structure is present.

Back to Top

 
 
Enzyme activity is low

For Roche Applied Science polymerases, 100% activity is guaranteed through the control date.

Suggestions:

  1. Check the control date. If necessary, obtain fresh enzyme
  2. Try "Hot start" technique to help retain activity through thermal cycling
Back to Top
 
 
Cycle number is too high

Most templates require 25-30 cycles.

Suggestion:

Cycle number should be based on starting concentration of template DNA.

If the number of target molecules in your sample is...

Then we recommend the following number of Cycles...

3 x 105 25-30
1.5 x 104 30-35
1.0 x 103 35-40
50 40-45
 
Back to Top
 
 
Nucleotides hydrolized
 

Suggestions:

Always store nucleotide stock solutions at a concentration of at least 10 mM, 100 mM is best. Significant hydrolysis occurs after storing at 1 mM for 2 months. Dissolve NTPs or dNTPs in water at an expected concentration of 10 mM. Using 0.05 M Tris-base and pH paper adjust the pH to 7.0. Dilute an aliquote of the neutralized NTP or dNTP appropriately and read the optical density at the wave lengths given in the following table. Calculate the actual concentration using the values for the extinction coefficients. Freeze away in small aliquots at -20°C.

Base

Wavelength Extinction coefficients for bases
e(M-1 cm-1)
A 259 1.54 x 104
T 260 7.4 x 103
G 253 1.37 x 104
C 271 9.1 x 103
U 262 1.0 x 104
 
The lithium and sodium salts have equivalent stability and work equally well in PCR, sequencing, and labeling applications. Lithium salts are more soluble in ethanol than sodium salts. Thus, removal of lithium salts by ethanol precipitation is more efficient than removal of sodium salts. Using lithium salt nucleotide preparations reduce salt-induced artifacts and increase the legibility of sequencing gels.
 
Back to Top
 
Nucleotide concentration is too high or unbalanced
 

The standard concentration is 20-200 µM of each nucleotide

Suggestions:

  1. Check the concentration of stock solutions of all nucleotides
  2. Double check the final concentrations of all nucleotides
Back to Top
 
 
Primer concentration is too low

Suggestion:

Adjust the primer concentration (0.1 - 1.0 µM of each primer is optimal)

Back to Top

 
 
Machine based error
Suggestions:
  • Calibrate the heating block and confirm actual block temperature
  • Run a diagnostic program for the particular machine - contact the machine manufacturer for specifics
Back to Top
 
 
Plateau effect
Possible causes of the Plateau Effect/solutions:
  1. Utilization of dNTP's - dNTP concentration should be 20-200 µM each
  2. End product pyrophosphate inhibition - reduce cycle number to 20-35, to reduce pyrophosphate formation
  3. Incomplete denaturation of product at high concentration - use stepwise cycling, increasing time of denaturation in later cycles
  4. Substrate excess in substrate/enzyme ratio - use stepwise cycling increasing extension time in later cycles or enzyme concentration
Back to Top
 
 
Evaporation

Evaporation can lead to higher concentrations of components, which may inhibit enzyme activity. The change in volume also leads to changes in the thermal profile inside the reaction tubes.

Suggestion: Use 100 µl mineral oil overlay/reaction

Back to Top

Roche Contact Information

Roche
Roche Diagnostics Corporation
P.O. Box 50414
9115 Hague Road
Indianapolis, IN 46250-0414 USA




Customer Service: 800-262-1640

Tech Support: 800-262-4911

Fax Number: 800-428-2883

Article Tools
  • Email a Colleague
  • Print
  • ShareThis
advertisement

Email Newsletter Sign-Up

Stay updated on the latest technologies and news with Biocompare's newsletters
(See samples here)












Select All

Loading

Loading