How Extensively Can Andor's Back-illuminated EMCCDs Be Used Before They Can No Longer Be Recalibrated (EMCAL™) To Factory EM Gain Settings?

How Extensively Can Andor's Back-illuminated EMCCDs Be Used Before They Can No Longer Be Recalibrated (EMCAL™) To Factory EM Gain Settings?

How extensively can Andor's back-illuminated EMCCDs be used before they can no longer be recalibrated (EMCAL) to factory EM gain settings?


One common concern associated with EM gain ageing phenomenon and the associated EMCAL recalibration fix, is that of longevity. The clock voltage setting which must be adjusted as part of the recalibration routine will eventually reach a maximum threshold value, after which further rescaling is not possible and EM gain will then fall off irreparably upon further extensive use of the camera. The question is, when is this likely to happen under typical use?

Andor have run extensive testing on the iXon+ DU-897 camera in order to project the operational lifetime of the 'gain register' (where signal amplification occurs on-chip) in back-illuminated EMCCDs, the conditions of testing described below:

Overall duration of test: ~14 months
Camera usage during this period: Continuous 24 hours per day, 7 days per week
Frame rate: 30 frames / sec
EM Gain setting: x1000
Photons / pixel / frame: 90
Number of pixels illuminated: ~ 200,000 (~ 75% of array)
EMCAL: Applied once per day


Results
The clock voltage required to maintain the EM gain calibration was measured once daily using Andor's EMCAL routine and the plot shown was generated by extrapolating from the equation derived from the gain ageing trend. This shows that under the aggresive test conditions employed, the EMCCD calibration would be expected to reach only ~ 65% of the available clock voltage scale after 7 years of continuous operation! As such, calibration over the ‘full EM gain range’ can be expected for significantly longer than 7 years.


Demanding test conditions
The combination of parameters employed in this test represents quite aggressive acquisition conditions. ~ 1.1 billion images were recorded during this period, with ~ 200,000 illuminated pixels per image, corresponding to ~ 220,000 billion pixels being amplified through the gain register with x1000 EM amplification per pixel.

Most users would not subject the camera to 24/7 continuous acquisition at 30 frames/sec. We also strongly recommend that, except for photon counting, the EM gain setting is limited to no more than x300 for the vast majority of applications, but x1000 was chosen here as a more rigorous test condition. Finally, it is quite rare that 75% off all pixels in the array will be subjected to uniform signal of this magnitude, as was imposed on the sensor here. In reality the light emitting species of typical user samples will project onto a much smaller fraction of pixels from frame to frame.



EMCAL does NOT accelerate EM gain ageing
It is important to recognize that the rate of ageing is not accelerated by routine application of Andor's EMCAL routine. The rate of ageing is determined by the illumination and EM gain conditions that the sensor is subject to through operation, irrespective of routine recalibration using EMCAL. If the 'previous' mechanism of EM gain recalibration were to be used, whereby the camera is shipped back to factory less frequently for manual readjustment of the clock voltage, the progress along the ageing curve would not be any different from that shown here (adjusting for the additional time that the camera would be out of action).


How often to run EMCAL?
This is a very difficult question to give a definitive answer to, because it depends very much on how extensively images are acquired and how 'aggressively' it is used in terms of EM Gain and light intensity across the sensor area. However, it is a reasonable rule of thumb to recommend that it is worth running EMCAL once per month. If the camera is in a multi-user facility, users may wish to run EMCAL more frequently for peace of mind - there is no technical downside to this; it will NOT accelerate the progress of the camera towards its EM gain end point.


Conclusion
For the vast majority of low-light applications and taking due care and attention to stay within recommended operating conditions, applying EMCAL as required, the gain ageing phenomenon is not considered to ever impose a restriction on the quantitative reliability of your Andor iXonEM+ camera.


back to top

Comments