Optimized Wavelength Scanning
of Fluorescent Proteins with the
SpectraMax Paradigm Platform and
TUNE Technology
The SpectraMax® Paradigm® Multi-Mode
Detection Platform with Tunable Wavelength
(TUNE) Detection Cartridge from Molecular
Devices combines the flexibility of wavelength
scanning with the high sensitivity of filters.
The TUNE Detection Cartridge uses a breakthrough
monochromator design that enables
researchers to easily optimize their assays over
a wide range of wavelengths with up to 10x
the sensitivity compared to other monochromator
systems. Excitation can be set from 360
nm to 790 nm, and emission can be set from
400 to 850 nm, in increments of 1 nm. TUNE
offers the additional functionality of time-resolved
fluorescence and luminescence detection
for an expanded range of assays.
A patent-pending Spectral Optimization Wizard
simplifies the workflow by automatically
calculating signal to background and determining
the optimal excitation and emission
wavelengths for any scanned fluorophores. A
heat map graphically displays calculated signal
to background results for scanned wavelength
pairs and allows users to read their assay using
automatically optimized wavelengths, providing
an 80-fold increase in sensitivity for EGFP
detection compared to simply using excitation
and emission peak wavelengths.
Excitation and emission spectral scanning
Individual excitation and emission spectral
scans were performed on two green fluorescent
protein variants. Recombinant wild-type GFP
(Clontech) and enhanced GFP (EGFP, BioVision)
were both diluted to 10 μg/mL in TE buffer,
pH 8. GFP samples (50 μL) were pipetted into a
black 384-well microplate for measurement of
spectral properties. Blank wells containing buffer
only were included for reference.
For individual excitation and emission spectral
scans, integration time was set to 140 ms. For
excitation scans, the emission wavelength was
set to 530 nm while scanning the excitation
spectra at 2-nm increments from 360 to 500 nm.
For emission scans, the excitation wavelength
was set to 440 nm while scanning the emission
spectra at 2-nm increments from 470 to 600 nm.
The excitation peak of EGFP was shown to be
red-shifted from 394 nm to 482 nm, while the
emission peaks remain similar at 506 nm (GFP)
and 512 nm (EGFP; Figure 1).
GFP and EGFP with TUNE (Figure 1)

Spectral Optimization Wizard
The Spectral Optimization Wizard in SoftMax®
Pro 6 Software was used to determine optimal wavelengths for detecting GFP and
EGFP. Users can select the excitation
and emission scanning ranges, as well
as the wavelength increment.
The Spectral Optimization Wizard scans
excitation and emission simultaneously
and calculates signal to background for
each excitation/emission wavelength
pair using the following formula: (Signal
– Background)/Background. Results are
displayed in a heat map where the excitation/
emission wavelength pair giving
the highest signal to background value is
highlighted with a crosshair icon. Results
for GFP are shown in Figure 2.
Optimized excitation and emission (Figure 2)

Sensitivity in detecting the GFP variants
is improved when using Wizard-optimized
wavelengths, as shown in Table 1.

For GFP, whose excitation and emission
peak wavelengths are separated by 112
nm, the increase in sensitivity afforded by
using the Wizard-optimized wavelengths
instead of excitation and emission peak
wavelengths is modest. However, for
EGFP, whose peak excitation and emission
wavelengths are separated by only 30 nm,
a greater than 80-fold increase in sensitivity
is provided by using the optimal
wavelengths calculated by the Spectral
Optimization Wizard. The optimal excitation
and emission wavelengths are set at
50 nm apart, so interference by scattered
excitation light is reduced.
Conclusion
The TUNE Spectral Optimization Wizard
reduces the time required for users to
optimize wavelengths for new fluorophores
by at least 50% by automatically
identifying an optimal wavelength pair
for any given fluorophore. Spanning a
range of 360-790 nm for excitation and 400‑850 nm for emission, SpectraMax
Paradigm with TUNE can detect all of the
most commonly used fluorophores with
up to 10x increased sensitivity over other
monochromator systems. TUNE-optimized
wavelengths offer the greatest benefit
for assays that use fluorophores with a
narrow Stokes’ shift and, as demonstrated
above, can result in an 80-fold increase in
sensitivity for EGFP detection. In addition
to fluorescence intensity, TUNE offers
time-resolved fluorescence and luminescence
detection modes for expanded assay
versatility.